Senseca ECO511 Ph Meter
Specificaties
- Merk
- Senseca
- Artikelnummer
- ECO511
- Categorie
- Meter
- Beschikbaarheid
- Op voorraad
Productbeschrijving
ECO 510 / ECO 511 - pH / Redox measuring device8Technical dataTechnical dataECO 510ECO 511pH0.00 .. 14.00 pH0.00 .. 14.00 pHRedox---1500 .. +1500 mV -1293 .. +1707 mVHTemperature---5.0 .. 105.0 °C (23.0 .. 221.0 °F)Measuring rangeAccuracy (at nom. temperature) pH± 0.02 pH ± 1 digit± 0.02 pH ± 1 digitRedox--± 0.1% FS ± 1 digitTemperature--± 0.3 °CConnection pH, (Redox)BNC connection for electrode BNC connection for electrodeTemperature--Banana 4 mm, Pt1000 (2-wire)Temperature compensation for pH-5 .. 105 °C (or 23 .. 221 °F)Input resistance pHapprox. 1012 OhmMeasuring cycleapprox. 2 measurements per secondDisplay3-line segment LCD, additional symbols, illuminated (white, luminous duration adjustable)Standard functionsMin/max/holdpH calibrationManual 1- or 2-point or automatic 2-point calibrationTemperature adjustmentOffset and gradient correctionHousingBreak-proof ABS housingProtection ratingIP65 / IP67 (only with electrodes identified as waterproof in the connected state for devices with BNC connection)Dimensions L*W*H108 * 54 * 28 mm, without BNC socket or kink protectionWeight (ECO 510 / ECO 511)~ 130 / 135 g incl. batteries, without electrode ~ 180 / 185 g incl. batteries and electrode GE 114Nominal temperature B-H86.0.1X.DK2-4.025 °C Page 27 of 32
Technische Documenten
Neem contact met ons op voor de beste prijs en snelle levering.